Effects of the Geometrical Dimensions of Applicator Antennas for Microwave Radiothermometry on Radiobrightness Temperature Measurement Results

We report here a study of the influences of the height and diameter of applicator antennas for microwave radiothermometry on radiobrightness temperature measurement results. The electric fields of three antennas based on numerical solution of Maxwell’s equations were modeled mathematically. The depth of the area whose temperature was measured was found to depend linearly on antenna diameter. An equation estimating the potential measurement depth of the antenna depending on its diameter for breast tissues was derived. A significant reduction in the height of the antenna (by a factor of six compared with traditional antennas) was found to have virtually no effect on the measured radiobrightness temperature in the projection of a malignant tumor, which is very important when designing flexible textile antennas.

Авторы
Vesnin S.G. 1 , Gudkov A.G.2 , Sedankin M.K. 1, 3, 4 , Leushin V.Y. 1 , Agasieva S.V. 1 , Porokhov I.O. 1 , Gudkov G.A.2 , Sidorova M.I. 1
Журнал
Издательство
Springer New York LLC
Номер выпуска
3
Язык
Английский
Страницы
190-194
Статус
Опубликовано
Том
57
Год
2023
Организации
  • 1 Peoples’ Friendship University of Russia
  • 2 OOO Hyperion Company
  • 3 A. I. Burnazyan Federal Medical Biological Center, Federal Medical Biological Agency of Russia
  • 4 MIREA Russian Technology University
Ключевые слова
biomedicine; general; biomaterials; engineering
Дата создания
01.07.2024
Дата изменения
01.07.2024
Постоянная ссылка
https://repository.rudn.ru/ru/records/article/record/109620/
Поделиться

Другие записи