Determination of Biapenem in Human Blood Plasma by High-Performance Liquid Chromatography with UV Detection

A technique for quantitative determination of biapenem in human blood plasma using HPLC with UV detection was developed to provide therapeutic drug monitoring in a hospital setting. Detection was made spectrophotometrically at 300 nm. The chromatographic procedure lasted 15 min. Precipitation with MeCN was used for sample preparation. The lower limit of quantification was 0.2 μg/mL. The methodology was validated for the main parameters. Linearity was observed over a wide concentration range from 0.5 to 100 μg/mL. The results made it possible to use the developed method for rapid determination of the biapenem content in patient blood plasma, which enabled personalization of the drug doses and increased the efficiency and rationality of antibacterial therapy.

Авторы
Номер выпуска
10
Язык
Английский
Страницы
1655-1660
Статус
Опубликовано
Том
57
Год
2024
Организации
  • 1 Peoples’ Friendship University of Russia
  • 2 Moscow City Health Department
Ключевые слова
biapenem; carbapenems; blood plasma; HPLC-UV; pharmacy; Pharmacology/toxicology; organic chemistry
Дата создания
01.07.2024
Дата изменения
01.07.2024
Постоянная ссылка
https://repository.rudn.ru/ru/records/article/record/111709/
Поделиться

Другие записи